We modified the software of Kostec¡¯s K-501A, a continuous ambient particulate monitoring instrument for use in this study. To calculate measurement uncertainty, we programmed the following four steps At first, establishing the model equation of PM concentration, at second, calculating each standard uncertainty, at third, calculating combined standard uncertainty, and finally, calcultaing expanded uncertainty of 95% confidence level. The uncertainty factors considered here are as follows. | |